Các trường hợp sửa chữa thiết bị y tế LINVATEC HD 4300 Arthroscope để sửa chữa
LINVATEC HD 4300 Arthroscope for repair, SN: 62**** Fault: Image with black shadow; eyepiece with scratches, inner optical parts broken. Replace objective lens, inner optical parts, clean and adjust optical system, seal with eyepiece. Image Quality Related Issues: 1.Fogging or Blurry Vision: This is ...
Các trường hợp sửa chữa thiết bị y tế của Olympus CF-H190L
Olympus CF-H190L Colonoscope for repair,SN:261****. Fault:Light bundle broken,biopsy tube worn,water air tube leakage,bending section deformation,wire helix tube worn. Solution:Replace the bending section,wire,helix tube,air tube and rubber.PS:Warranty for replace parts only. Repair time:7 working ...
Olympus A4674A Ống soi buồng tử cung hấp tiệt trùng để sửa chữa
Olympus A4674A Autoclavable Hysteroscope for repair ► Brief Description Fault: 1. Image blur and black shadow. 2. Rod lens broken. 3. Outer tube bending and transformation. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the ...
Khóa học chứng nhận sửa chữa thiết bị y tế về đào tạo nội soi linh hoạt cho khách hàng Hàn Quốc
The courses we organize and their contents are as follows. COURSE: Flexible Endoscope Repair Training Course Brands: OLYMPUS, FUJINON Gastroscope, Colonoscope, Duodenoscope, Bronchoscope, Fiberscope, Enteroscope, Hysteroscope Duration: 5 Days / 4 hours per day (20 hours total) Course Schedule: ...
C5-2 Máy thăm dò cong để sửa chữa các bộ phận thiết bị y tế
C5-2 Convex Probe For Repair ► Brief Description Fault: Strain relief damage, cable broken. Solution: Replace strain relief, repair cable. ►Problem & Solutoin Model C5-2 Convex Fault Strain relief damage, cable broken Solution Replace strain relief, repair cable ► Common Issues with C5-2 Probes Dead ...
GE 4C-RS thăm dò bụng để sửa chữa
GE 4C-RS abdominal probe for repair ► Brief Description Fault: Lens broken, cable broken, crystal bad. Solution: Replace probe head, repair cable. ►Problem & Solutoin Bran GE Model 4C-RS Fault Lens broken, cable broken, crystal bad Solution Replace probe head, repair cable ► Common issues with 4C-RS ...
S5-2 Đầu dò tim mạch để sửa chữa
S5-2 Cardiac probe for repair Fault: Lens problem, cable broken, without strain relief. Solution: Replace probe lens, repair cable, replace strain relief. ►Problem & Solutoin Model S5-2 Fault Lens problem, cable broken, without strain relief Solution Replace probe lens, repair cable, replace strain ...
Karl Storz 27005BA Cystoscope for repair
Karl Storz 27005BA Cystoscope for repair ► Brief Description Fault: 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens ...
Sonoscape C362 Chuẩn bị sửa chữa
Sonoscape C362 Abdominal probe for repair ► Brief Description Fault: Cystal problem. Solution: Replace the Crystal. ►Problem & Solutoin Brand Sonoscape Model C362 Fault Cystal problem Solution Replace the crystal ► Common issues with C362 Probe Image Dropouts (Black Lines or Streaks) Damaged Lens ...